Abstract

Development and Validation of a HPLC Method for Simultaneous Estimation of Lamivudine and Zidovudine in Tablet Dosage Forms

A simple, precise and accurate HPLC method has been developed for the simultaneous estimation of lamivudine and zidovudine in tablet formulations. The chromatographic separation was achieved on a Hypersil SS C18 (250 mm × 4.6 mm, 5μm) analytical column. Methanol was used as the mobile phase, at a flow rate of 1 ml/min and detector wavelength at 270 nm. The validation of the method was accomplished for specificity, linearity, accuracy, precision, limit of detection (LOD), limit of quantification (LOQ) and robustness. The linear dynamic ranges were from 0.1–2.0μg/ml for lamivudine and zidovudine. The percentage recovery obtained for lamivudine and zidovudine were 100.36 and 100.46 %, respectively. Limit of detection and quantification for lamivudine were 0.042 and 0.12 μg/ml, for zidovudine 0.039 and 0.12 μg/ml, respectively. The developed method can be used for routine quality control analysis of lamivudine and zidovudine in combination in tablet formulation.


Author(s): Ravi Prakash Mahor1 , Versha Parcha2, Yogendra Singh1, Rajiv Sharma3 and Anil Bhandari

Abstract | PDF

Share This Article
Awards Nomination 17+ Million Readerbase
Google Scholar citation report
Citations : 5682

Der Chemica Sinica received 5682 citations as per Google Scholar report

Der Chemica Sinica peer review process verified at publons
Abstracted/Indexed in
  • Google Scholar
  • Open J Gate
  • Genamics JournalSeek
  • China National Knowledge Infrastructure (CNKI)
  • Directory of Research Journal Indexing (DRJI)
  • Publons
  • MIAR
  • International Committee of Medical Journal Editors (ICMJE)
  • Serials Union Catalogue (SUNCAT)
  • Geneva Foundation for Medical Education and Research
  • Secret Search Engine Labs
  • Euro Pub
  • CAS (Chemical Abstracting Services)
  • University of Barcelona

View More »

Flyer image
T�rkiye'nin Yeni Nesil Bahis Adresi Matadorbet giris Adresi G�ncellenmistir!